Product Identification

  • Manufacturer: Gefran.

  • Model: CU-K1M 2130X000X00.

  • Ordering Code: F005165.

  • Product Series: CU.

  • Product Type: Compact Compression Load Cell.

  • Measurement Principle: Strain gauge bridge.

  • Load Direction: Compression.

Technical Specifications

  • Nominal Capacity: 1,000 kg.

  • Accuracy Class: 0.2%.

  • Nominal Full-Scale Output: 2 mV/V.

  • Output Tolerance at Nominal Load: Less than ±0.2% FSO.

  • Combined Error: Less than ±0.2% FSO.

  • Creep after 30 Minutes at Nominal Load: Less than ±0.06% FSO.

  • Zero Balance: Less than ±1% FSO.

  • Sensitivity Thermal Drift: Less than ±0.01% FSO/°C.

  • Zero Thermal Drift: Less than ±0.01% FSO/°C.

  • Nominal Input Resistance: 350 Ω.

  • Nominal Output Resistance: 350 Ω.

  • Insulation Resistance: Greater than 10 GΩ.

  • Nominal Excitation Voltage: 10 V.

  • Maximum Excitation Voltage: 15 V.

  • Compensated Temperature Range: -10°C to +50°C.

  • Maximum Operating Temperature Range: -20°C to +60°C.

  • Storage Temperature Range: -30°C to +80°C.

  • Permitted Static Load: 130% of nominal load.

  • Permitted Dynamic Load: 100% of nominal load.

  • Maximum Applicable Load: 150% of nominal load.

  • Rupture Load: Greater than 300% of nominal load.

  • Maximum Elastic Deformation at Nominal Load: Less than 0.2 mm.

  • Protection Rating: IP67 (DIN 40050).

  • Elastic Element Material: Stainless steel.

  • Outside Diameter: 50 mm.

  • Electrical Connection: 4 × 0.25 mm² screened cable.

  • Cable Length: 5 m.

 
gefran.sk

Electrical Connections

  • Signal +.

  • Signal -.

  • Excitation +.

  • Excitation -.

  • Shield.

The cable shield is isolated from the load cell body. Gefran recommends grounding the shield at the measuring instrument end.

Mechanical Features

  • One-piece stainless steel body.

  • Weld-free construction.

  • Compact 50mm outside diameter.

  • Four M5 mounting holes.

  • Compression loading direction as specified in the manufacturer's mechanical drawing.

Typical Applications

  • Industrial compression force measurement.

  • Industrial weighing systems.

  • Machine load monitoring.

  • Industrial press force sensing.

  • Static and dynamic compression testing.

  • Corrosion-resistant force measurement applications.

Important Technical Notes

The CU-K1M is intended for compression measurement. It must not be treated as a tension/compression load cell without explicit manufacturer confirmation.

The load cell requires compatible signal conditioning or measurement equipment and is not a standalone 4–20mA transmitter.