Model
|
HR-1100
|
Display
|
LCD
|
Waterproof specifications
|
Equivalent to IPX5
|
Input connector
|
ASP
|
Input type
|
Thermocouple Input: Type E, K
|
Channel
|
1ch
|
Signal source resistance
|
MAX.1k ohm
|
Measurement Range
|
1°C Resolution
|
E
|
-200 to 800°C
|
K
|
-200 to 1370°C
|
0.1°C Resolution
|
E
|
-104.9 to 504.9°C
(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.)
|
K
|
-104.9 to 504.9°C
(Switches automatically to 1°C resolution range when messured temperatures fall outside the 0.1°C resolution range.)
|
Measurement Accuracy
|
1°C Resolution
|
0°C to
|
± (0.1% of reading +1°C)
|
to 0°C
|
± (0.5% of reading +1°C)
|
0.1°C Resolution
|
0°C to
|
± (0.05% of reading + 0.2°C)
|
to 0°C
|
± (0.15% of reading + 0.2°C)
|
Reference junction compensation accuracy
|
±0.2°C at 25°C±10°C
|
Temperature Coefficient*1
|
±0.02×Δt°C (Add the value obtained by multiplying the excess temp.Δt[°C] by a coefficient to the total accuracy. e.g. ±0.3°C is added at room temp. of 0°C or 50°C)
|
Operation Environment
|
0 to 50°C, 0 to 80%RH (Non condensing)
|
Storage Environment
|
-20 to 50°C, 0 to 85%RH (Non condensing)
|
Battery life
|
approx.900h
|
Power supply(Dry battery)
|
4 AA Alkaline batteries
|
Power supply(AC power)
|
-
|
Sampling rate
|
approx.200ms
|
Linearlizer
|
Digital Linearizer method (Compliant with JIS C1602-2015)
|
Dimensions(mm)
|
approx. 82(w)×166(H)×36(D) [Protruded parts excluded]
|
Weight
|
approx. 350g [Include Batteries]
|
Supplied accessories
|
Manual,Test report,Soft case,
|
Hand strap,4 AA Alkaline batteries
|
Compatible standards
|
EMC:EN61326-1:2013, EN 61326-2-1:2013 classA Table2( Industrial)
RoHS:IEC EN 63000:2018
|