HAUBER-Elektronik HE103.01.16.01.62.00.200 – Vibration Monitoring Unit, 0–16 mm/s RMS, 4–20 mA, Duplex 1.4462, 20 m Cable
Price: Contact
Brand: HAUBER-Elektronik
Category: Cảm biến công nghiệp
Supplier: AnhNghiSon
Origin: Germany
1. Product Identification
Manufacturer: HAUBER-Elektronik GmbH.
Model: HE103.01.16.01.62.00.200.
Article Number: 18038.
GTIN: 4255921340704.
Series: HE103.
Product Type: Vibration Monitoring Unit.
Measured Quantity: Vibration velocity RMS.
Monitoring Application: Absolute bearing vibration measurement.
2. Measurement Specifications
Measuring Range: 0–16 mm/s RMS.
Frequency Range: 1–1,000 Hz.
Accuracy: ±10% according to DIN ISO 2954.
Calibration Point: 159.2 Hz at 90% of the measuring range.
Corresponding Calibration Value: 14.4 mm/s RMS.
Cross Sensitivity: Less than 5%.
Maximum Acceleration: ±16.5 g.
Measuring Direction: Along the mounting axis.
3. Extended RMS Averaging
Averaging Time: 60 seconds.
Signal Evaluation: Exponentially time-weighted RMS.
Reference Settling Response: 99% after approximately 300 seconds.
Function: Reduction of the influence of stochastic environmental disturbances.
Application: Stable vibration signals for long-term condition monitoring.
The averaging time does not represent a fixed 60-second analogue output update interval.
4. Analogue Output Specifications
Output Type: Analogue DC current output.
Output Signal: 4–20 mA.
Output Characteristic: Proportional to the vibration measuring range.
Nominal Lower Scale Value: 4 mA at 0 mm/s RMS.
Nominal Mid-Scale Value: 12 mA at 8 mm/s RMS.
Nominal Upper Scale Value: 20 mA at 16 mm/s RMS.
Maximum Published Output Load: 500 Ω.
Number of Outputs: One.
Transmitter Design: Two-wire system.
5. Electrical Specifications
Supply Voltage: 10–30 VDC.
Reference Nominal Supply Voltage: 24 VDC.
Maximum Current Consumption: 25 mA.
Power and Signal Connection: Two-wire current loop.
Integrated Dual Alarm Outputs: Not provided as in the HE200 series.
External alarm functions may be implemented through compatible industrial monitoring systems.
6. Mechanical Specifications
Housing Material: Duplex stainless steel 1.4462.
Reference Material Designation: Duplex 2205.
Cable Protection: Mounted metal protection hose.
Mounting Thread: External M8 × 8 mm.
Thread Pitch: 1.25 mm.
Sensor Width Across Flats: 24 mm.
Sensor Tightening Torque: 8 Nm.
Installation Orientation: Horizontal or vertical.
Measuring Axis: Along the mounting axis.
Published Configuration Weight: Approximately 4,062.5 g.
The published weight should not be interpreted as the weight of the bare sensing head.
7. Integrated Cable Specifications
Cable Length: 20 metres.
Cable Type: 12YC11Y.
Cable Construction: Four conductors, 0.34 mm² each.
Conductor Material: Stranded copper.
Core Insulation: TPE-E (12Y).
Outer Sheath Material: PUR.
Outer Sheath Diameter: 6.0 ±0.2 mm.
Fixed Installation Cable Temperature: −40 to +80°C.
Moving Cable Temperature: −30 to +80°C.
Minimum Fixed Installation Bending Radius: 30 mm.
Minimum Moving Cable Bending Radius: 60 mm.
Halogen-Free Construction: Yes.
Flame Resistance Reference: UL 1582 Section 1061.
Cable specifications must not automatically be applied to the complete metal protection hose assembly.
8. Connection Identification
Brown Conductor: L+, 10–30 VDC.
Blue Conductor: 4–20 mA current-loop signal.
White Conductor: Not connected.
Black Conductor: Not connected.
Connection Type: Integrated four-core cable with two-wire transmitter operation.
Hazardous area wiring and commissioning must comply with the manufacturer's applicable safety documentation.
9. Environmental and Reliability Specifications
Maximum Published Relative Humidity: 100%.
Outdoor Application: Suitable according to manufacturer product data.
Design Lifetime: 10 years.
MTTF: 399 years.
MTTF is a statistical reliability parameter and does not represent a guaranteed service life or warranty period.
10. Explosion Protection
Manufacturer-Listed Approvals: ATEX, IECEx, CN-Ex and KCs.
Gas Hazardous Areas: Zone 1 and Zone 2.
Dust Hazardous Areas: Zone 21 and Zone 22.
Published Gas Protection Marking: Ex II 2 G Ex d IIC T4 Gb.
Published Dust Protection Marking: Ex II 2 D Ex tb IIIC T120°C Db.
Gas Group: IIC.
Gas Temperature Class: T4.
Gas Equipment Protection Level: Gb.
Dust Group: IIIC.
Published Maximum Surface Temperature for Dust Protection: T120°C.
Dust Equipment Protection Level: Db.
The final equipment marking and certification scope must be verified against the supplied instrument and applicable certificates.
11. Reference Standards
DIN ISO 10816 – Evaluation of machine vibration.
ISO 20816 – Measurement and evaluation of machine vibration.
DIN ISO 2954 – Requirements for vibration severity measuring instruments.
ATEX – Equipment for potentially explosive atmospheres.
IECEx – International explosion protection certification system.
CN-Ex – Chinese explosion protection certification.
KCs – Korean certification requirements.
Additional manufacturer-listed conformity markings include CE, IEC and cULus Ordinary Locations.
12. Key Features
Continuous bearing vibration monitoring.
Vibration velocity range of 0–16 mm/s RMS.
Frequency range of 1–1,000 Hz.
Extended 60-second RMS averaging.
Stable analogue measurement signals.
Two-wire 4–20 mA transmitter.
Duplex stainless steel 1.4462 construction.
Integrated 20-metre cable with metal protection hose.
Hazardous area configuration.
Suitable for long-term industrial condition monitoring.
13. Industrial Applications
Electric motor vibration monitoring.
Industrial pump condition monitoring.
Fan and blower vibration measurement.
Generator bearing condition monitoring.
Turbine vibration monitoring.
Chemical processing equipment.
Oil and gas rotating machinery.
Marine industrial machinery.
Water treatment equipment.
Power generation facilities.
Industrial maintenance and condition monitoring systems.
14. Application Limitations
The extended averaging response is intended for stable vibration monitoring rather than rapid transient detection.
The maximum acceleration capability must be considered when evaluating the available vibration velocity operating range across the complete frequency band.
Functional safety certification according to SIL 2 or PL-d must not be assumed solely from certifications applicable to other HAUBER product families.